Simulation-Based Evaluation of Signal-Quality Gating for Wearable Respiratory Monitoring During Procedural Sedation

Phan Minh Duc

ABSTRACT
Background: Motion artifacts and poor attachment can compromise chest-motion respiratory monitoring. This study evaluated the effect of a signal-quality gate on respiratory-rate estimates and motion-pause alerts.

Methods: The benchmark comprised 240 synthetic traces across six scenarios (40 per scenario), with two programmed pauses per trace (80 per scenario and 480 overall). Both algorithms processed identical traces and used the same rate estimator.

Results: Baseline and gated coverage were 89.7% and 46.2%, respectively. Mean absolute error (MAE) was 0.14 and 0.07 breaths/min on the respective sets of available estimates. All 1,830 shared rate estimates were identical. Pause sensitivity fell from 438/480 (91.25%) to 400/480 (83.33%), and false alerts fell from 100 to 20 over 12 hours (8.33 to 1.67 per hour). All 20 gated false alerts occurred in the shallow-motion scenario. During movement bursts, gating retained 16.0% of eligible rate windows and detected 0/80 pauses.

Conclusions: The gate reduced false alerts by selectively withholding output, with lower rate coverage and pause sensitivity. The findings apply to the six synthetic scenarios and identify conditions for subsequent physical bench testing.